Laser Flash Photolysis System
TSP-1000/TSP-1000M
PDF

TSP-1000 is a laser flash photolysis system for the measurement of transient absorbance. There are many uses in the analysis of the elementary processes of photochemical reactions. With a nanosecond pulsed laser as the exciting light source, this system measures transient UV-VIS absorption in a broad range from nanosecond to millisecond.
Data are acquired by the combination of a photomultiplier detector and a digital oscilloscope in the single-wavelength monochromator system, and by the combination of a photodiode array detector and a high-speed AD converter in the multi-channel spectrophotometer system. In both systems, time resolution up is available to 10 nsec.
Exciting Light Source (pulsed YAG laser) | |
---|---|
Pulse Width | 4 ~ 6 nsec |
Energy (TSP-1000) | 200 mJ @ 532 nm, 60 mJ @ 355 nm |
Energy (TSP-1000M) | 25 mJ @ 532 nm, 8 mJ @ 355 nm |
Repetition Rate | 10 Hz (single pulse available) |
Monitoring Light Source | |
Lamp | 150 W Xe arc lamp of high stability |
Lamp House | Air-cooling type |
Light Intensity Adjustment | Iris function with a space to insert an optical filter |
Light Guide for Light Path | |
Material | Fiber optics made of quartz (transmittable in the range of 200 nm ~ 1000 nm) |
Length | 1 m |
Sample Cell Chamber | |
Temperature Range | 5°C ~ 60°C |
Temperature Control | By a bath circulator |
Cell | 10 mm2 quartz cell (transparent in 4 sides) |
Spaces for Optical Filters | At the exciting side and the detection side |
Data Processor | |
OS | Windows® 7 or later |
Control Interface | DIO (PCI), GPIB (PCI) |
Software | Averaging, Overlaying, Converting of wavelength axis and time axis, Non-linear least square fitting, Data storing in text format |
Spectrometer MF200 | |
---|---|
Optical Alignment | Czerny-Turner grating |
Focal Distance | 200 mm |
Speed | F/3.5 |
Linear Dispersion | 8.5 nm/mm |
Slit Width | 0.1 mm, 0.2 mm, 0.5 mm |
Monochromatic Detector | |
Detector | Photomultiplier tube |
Response | 5 nsec or less |
High-Voltage Power Supply | 0 ~ 1100 V adjustable |
Control Circuit | Wavelength scanning circuit |
Digital Oscilloscope | High-voltage feedback circuit, Sampling rate: 1.25 GS/sec max, Vertical sensitivity: 1 mV/div ~ 10 V/div, Time axis range: 40 nsec/div ~ 10 sec/div |
Spectrometer MD 308 | |
---|---|
Optical Alignment |
Czerny-Turner, grating |
Gratings | 3 different gratings automatically changeable |
Focal Length |
300 mm |
Speed | F/4 |
Linear Dispersion |
5.4 nm/mm (when using the 600 g/mm grating) |
Slit Width |
10 mm ~ 3 mm continuously changeable |
Multi-Channel Detector | |
Detector | Photodiode array 1024ch with a gated image intensifier |
Time Resolution |
5nsec ~ 10μsec |
Image Intensifier Gain |
Adjustable |
Synchronization Control |
Pulse generator with GPIB interface |
Wavelength Width |
100 nm, 200 nm, 400 nm (400 nm when using MD 200) |
Optional Accessories
- Exciting light source 266 nm
- OPO wavelength-changeable laser
- Low-temperature cell chamber (-80°C ~ +100°C)
- Stopped-flow mixer (for flow flash measurement)
- Bath Circulator
- Optical filters